No open-access location is known for this record yet.
No open-access location is known for this record yet.
We describe the dynamic evaluation of multi-component force–moment transducers. Two of these transducers, which can measure three force components, F x , F y and F z , and three moment components, M x , M y and M z , were developed. One has a solid cylindrical sensing element and the other has a hollow cylindrical shape. They were calibrated statically with a force–moment calibration machine in the Physikalisch-Technische Bundesanstalt (PTB), Germany. We evaluated the dynamic characteristics of the force–moment transducers by using a shaker system and a multi-channel dynamic analyser system. The frequency range of the normal force component F z is very wide up to 1 kHz, the test frequency in this study. However, the frequency ranges for the transverse force components and moment components are restricted to about 200 Hz and 130 Hz respectively. The column-type force–moment transducer has a high stiffness and so its frequency range is wider than that of a binocular-type force–moment transducer.
Loading…
Stored only when both papers are in the corpus — 12 citations exist upstream.
Papers that cite the same references — the relational fallback while embeddings are still backfilling.
Imported from OpenAlex (CC0) — nothing here is generated.