This protocol was developed to estimate the uncertainty of measurements in the sector field inductively coupled plasma mass spectrometry analysis of As, Co, Cr, Mn, Mo, Ni, Sn and V in human serum by utilizing in-house validation data. The approach is outlined in the Eurachem/Citac Guide and other relevant publications. The results generated were the followings: (i) coefficients of regression >0.9995 over two orders of magnitude of metal concentration; (ii) method quantification limits between 0.05 µg/L (Cr, Mn) and 0.49 µg/L (As); (iii) mean trueness checked against a certified reference material between 95.4% (As) and 107.7% (Ni); (iv) repeatability better than 10.2% over the range 0.1-2.0 µg/L; (v) reproducibility better than 12.0% over the range 0.1-2.0 µg/L; and (vi) expanded uncertainty budget comprised between 14.7% (Mn) and 27.9% (Cr) over the range 0.1-2.0 µg/L.