No open-access location is known for this record yet.
No open-access location is known for this record yet.
The aim of this comparison is to evaluate the gravimetry and purity verification of gas mixtures prepared by introducing a liquid into a gas cylinder. This type of preparation is well known from, e.g., ethanol in nitrogen/air (CCQM-K4), and n -hexane in synthetic natural gas ( CCQM-K16 ). The choice for n -hexane in methane is based on the relevance of natural gas primary standard mixtures and higher hydrocarbons in synthetic natural gas (for, e.g., dew point calibrations). Furthermore, n -hexane is relatively straightforward to analyse using a GC/FID (gas chromatograph equipped with a flame ionisation detector) with good repeatability. Eight laboratories participated. After the data evaluation, the result of one laboratory was eliminated to arrive at a consistent subset. The quality of the fitted line was not overly good, leading to small discrepancies in the degrees-of-equivalence calculated for three other laboratories. Only four laboratories are consistent with the KCRV. Main text. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/ . The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
Loading…
Stored only when both papers are in the corpus — 8 citations exist upstream.
Imported from OpenAlex (CC0) — nothing here is generated.