No open-access location is known for this record yet.
No open-access location is known for this record yet.
A description is given of the results of a Key Comparison of primary free-field standards for underwater acoustics at frequencies from 1 to 500kHz. This is the first such Key Comparison exercise in the field of underwater acoustic calibration and measurement. Laboratories from UK, Germany, USA, Russia, China, Canada, and South Africa participated by calibrating three reference hydrophones, with project coordination provided by the National Physical Laboratory, UK. The agreement between the results obtained from the comparison was generally encouraging, with the calibration values reported by the laboratories agreeing within quoted uncertainties over the majority of the frequency range, and the results generally lying within a ±0.5-dB band for frequencies up to 300kHz. A discussion is given of the general sources of uncertainties in the calibrations, in particular those which are thought to have contributed to the differences in the results between laboratories. The results of the participants have been used to estimate the equivalence of national measurement standards within this field.
Loading…
Stored only when both papers are in the corpus — 37 citations exist upstream.
Papers that cite the same references — the relational fallback while embeddings are still backfilling.
Imported from OpenAlex (CC0) — nothing here is generated.