No open-access location is known for this record yet.
No open-access location is known for this record yet.
A comparison of two standard humidity generators (two-temperature (2-T) and one-temperature (1-T) generators) that are developed by the National Institute for Standards (NIS) has been performed using a transfer standard chilled-mirror hygrometer and measurement procedures to realize dew-point temperature Td in the range from −50 °C to 0 °C. The main objective of this comparison was to compare the realizations of dew-point temperature and to establish the level of consistency between the two generators. For a level of consistency between two measurements, it is expressed by the difference between the measured values, m1 − m2, and the expanded pair uncertainty of this difference Up [1]. The comparison measurements revealed dew-point temperature differences of 0.02 °C and 0.07 °C with expanded pair uncertainties of ±0.09 °C and ±0.15 °C.
Loading…
Stored only when both papers are in the corpus — 22 citations exist upstream.
Imported from OpenAlex (CC0) — nothing here is generated.